The Ametek T-960 and T-970 pneumatic hand pumps are designed for generating precise pressure to verify, adjust, and calibrate mechanical and electronic pressure measuring instruments. The T-960 covers 0 to 30 psi (2 bar), while the T-970 extends to 580 psi (40 bar).

Both models feature an extended range vernier adjustment valve for precise pressure control and a comfortable pistol grip handle for ergonomic operation. The pump includes both reference and test pressure ports, with the reference instrument threading directly to the top of the pump via a quick connector.

These pumps are compatible with DPC-500, APC, CPC, and IPI calibrators. A key design advantage is that the pump and indicator are separate systems, allowing field replacement or local recalibration without returning the complete system to the factory. Recalibration can be performed locally using an accurate pressure reference; only the indicator needs to be returned for factory calibration if required.

JOFRA pressure calibrators are available as individual indicators or as complete ready-to-use systems in a carrying case, including pump, hose, fitting set, batteries, test leads, and NIST-traceable calibration certificate.

Features

  • Two models: T-960 (0-30 psi) and T-970 (0-580 psi)
  • Extended range vernier adjustment for precise pressure control
  • Comfortable pistol grip handle
  • Separate pump and indicator – easy maintenance and local recalibration
  • Compatible with DPC-500, APC, CPC, and IPI calibrators
  • Reference port with quick connector for instrument mounting
  • Available as pump only or complete calibration system
  • Compact design: 216 x 121 x 62 mm
  • Media: clean, dry air
Ametek

Pneumatic Pressure Pumps - 580 psi & 30 psi Ranges | Ametek T-970 / T-960

Pistol grip design with extended vernier for precise control
MPN/Series: T-960 / T-970

Product Specifications

Manufacturer:Ametek Calibration (Jofra)

Product Number:T-960 / T-970

Type:Pneumatic Pressure Hand Pump

T-960 Range:0 to 30 psi (0 to 2 bar)

T-970 Range:0 to 580 psi (0 to 40 bar)

Media:Clean, dry air

Fine Adjustment:Extended range vernier valve

Pressure Connections:Reference port (top), test port (hose)

Handle:Comfortable pistol grip

Dimensions:216 x 121 x 62 mm

System Compatibility:DPC-500, APC, CPC, IPI calibrators

Design Feature:Pump and indicator separate – local recalibration possible

Documents: Product Manual , Product Specs