Product Specifications
Manufacturer: Rosemount
Product Number: Model 150
Product Type: Insertion/Submersion Contacting Conductivity Sensor
Available Cell Constants: 0.1, 0.5, 1.0/cm
Conductivity Range: 0-10 to 0-10,000 µS (compatible with Rosemount 54eC/81C/1054BC/1181C/2081C/3081C analyzers/transmitters)
Temperature Compensation Range: 32 to 250°F (0 to 121°C)
Pressure Range: 50 psi to 150 psi
Process Connection: 1 inch MNPT (front and rear)
Housing Material: Tefzel
Electrode Material: 316 stainless steel
Insulator/Seal Materials: Tefzel and Viton
Standard Cable: Integral 10 foot cable
Mounting Options: Insertion into process line/tank, submersible installation, flow-through installation (with 1 inch CPVC tee accessory)
Primary Applications: Boiler Blowdown, Open Recirculating Cooling Water, Set-Point Detection with Broad Control Range
Documents: Product Manual , Product Specs
Product Description
The Rosemount Model 150 Insertion/Submersion Sensor is available in sensor constants of 0.1, 0.5 or 1.0 for conductivity ranges from 0-10 to 0-10,000 μS when used with Models 54eC, 81C, 1054BC, 1181C, 2081C, and 3081C analyzers/transmitters. For optimal loop performance, connect the Model 150 sensor to instrument Models 54eC or 1054BC.
The Tefzel housing has both rear-facing and forward-facing, 1 inch male pipe threads. The sensor can be threaded into a process line or the side of a tank, or be attached to a length of pipe for submersible installation. The Model 150 is available with a 1 inch CPVC tee (Code 16) for flow-through installation. An integral 10 foot cable is supplied for connection to the analyzer/transmitter or junction box.
This sensor is most appropriately used as a set-point detector in applications with a fairly broad control range. This sensor is a good choice for boiler blowdown and open recirculating cooling water. For applications that require high accuracy over a wide range of conductivity, a sensor from the ENDURANCE™ family (Model 400 series) or the toroidal family (Model 200 series) would be the ideal choice.
Features
-
Temperature Compensation:
- 32 to 250°F (0 to 121°C)
-
Pressure Range:
- 50 psi to 150 psi
-
Process Connection:
- 1 inch MNPT (front and rear)
-
Wetted Materials:
- Electrodes; 316 stainless steel
- Insulators; Tefzel and Viton
Resources & Documentation
Insertion/Submersion Conductivity Sensor - Boiler Blowdown, Cooling Water | Rosemount Model 150
Product Specifications
Manufacturer: Rosemount
Product Number: Model 150
Product Type: Insertion/Submersion Contacting Conductivity Sensor
Available Cell Constants: 0.1, 0.5, 1.0/cm
Conductivity Range: 0-10 to 0-10,000 µS (compatible with Rosemount 54eC/81C/1054BC/1181C/2081C/3081C analyzers/transmitters)
Temperature Compensation Range: 32 to 250°F (0 to 121°C)
Pressure Range: 50 psi to 150 psi
Process Connection: 1 inch MNPT (front and rear)
Housing Material: Tefzel
Electrode Material: 316 stainless steel
Insulator/Seal Materials: Tefzel and Viton
Standard Cable: Integral 10 foot cable
Mounting Options: Insertion into process line/tank, submersible installation, flow-through installation (with 1 inch CPVC tee accessory)
Primary Applications: Boiler Blowdown, Open Recirculating Cooling Water, Set-Point Detection with Broad Control Range
Documents: Product Manual , Product Specs