Product Specifications
Manufacturer:Isothermal Technology Limited (Isotech)
Product Number:999 Medusa R
Type:Primary Blackbody Calibration Source
Temperature Range:30°C to 550°C
Emissivity:>0.995
Stability:±0.05°C
Accuracy:±0.05°C (with fixed point cells)
Display Resolution:0.01°C (30-99.99°C); 0.1°C (100-550°C)
Cavity Size:45 mm diameter x 285 mm depth
Heating Time:45 minutes to 550°C
Stabilization Time:45 minutes
Controller Stability (5 min):±0.2°C
Cavity Uniformity:±1°C (100-500°C at 48mm target); ±4°C along 200mm length
Fixed Point Compatibility:Indium (156.5985°C), Tin (231.928°C), Zinc (419.527°C); large fixed point cells available
Power Supply:100-130V or 208-240V AC, 50/60 Hz
Power Consumption:1000 W typical
Dimensions (HxWxD):480 mm x 425 mm x 260 mm
Weight:17 kg
PC Interface:Included with Windows software
Calibration Certificate:UKAS or NPL traceable calibration available
Included Probe:935-14-95 PRT (optional)
Optional Fixed Point Cells:998-06-00A (Indium), 998-06-00B (Tin), 998-06-00C (Zinc), 984-00-00 Gas Flow System
Product Description
The Isotech 999 Medusa R is a primary blackbody source specifically designed for calibrating radiation pyrometers and infrared thermometers in standards laboratories requiring the highest levels of measurement accuracy. With a temperature range of 30°C to 550°C, it serves as an optimal radiation source for both comparison calibrations and fixed point realizations.
The instrument is engineered to accommodate ITS-90 blackbody fixed point cells, including indium (156.5985°C), tin (231.928°C), and zinc (419.527°C). When used with these 99.9999% pure metal cells, the Medusa R can achieve accuracies of ±0.05°C, making it suitable for primary standard applications. The 45 mm diameter by 285 mm deep cavity provides excellent thermal uniformity with emissivity greater than 0.995.
The digital temperature controller allows precise setting of cavity temperature with resolution of 0.01°C below 100°C and 0.1°C above. Warm-up time to 550°C is approximately 45 minutes, with similar stabilization time. Temperature variation over a 5-minute period is ±0.2°C as indicated on the controller, with similar variations detected by radiation thermometers viewing the cavity.
Uniformity testing reveals maximum temperature differences of ±1°C across the 48 mm target area for set points between 100°C and 500°C. Along the inner 200 mm of the cavity, maximum differences of ±4°C were measured at 400°C and 500°C using a fiber-optic radiation probe.
The Medusa R includes PC interface and Windows software as standard. It is available with UKAS or NPL traceable calibration certificates for the built-in indicator and supplied probe (935-14-95 PRT). An optional gas flow system (984-00-00) is available for fixed point operation.
Applications include primary laboratory calibration of radiation thermometers, ITS-90 fixed point realization, and high-accuracy infrared thermometer verification in national metrology institutes and accredited calibration facilities.
Features
- Temperature range: 30°C to 550°C
- Emissivity >0.995
- Stability ±0.05°C; Accuracy ±0.05°C (with fixed points)
- 45 mm diameter x 285 mm depth cavity
- Designed for ITS-90 fixed point cells (indium, tin, zinc)
- Freezes and melts blackbody fixed-point cells
- Suitable for comparison calibration
- 45-minute warm-up to 550°C
- PC interface and Windows software included
- Optional large fixed point cells for primary standards
- UKAS/NPL traceable calibration available
- Dimensions: 480 x 425 x 260 mm; Weight: 17 kg
Resources & Documentation
Primary Blackbody Source - 30-550°C, Fixed Point Cells | Isotech 999 Medusa R
Product Specifications
Manufacturer:Isothermal Technology Limited (Isotech)
Product Number:999 Medusa R
Type:Primary Blackbody Calibration Source
Temperature Range:30°C to 550°C
Emissivity:>0.995
Stability:±0.05°C
Accuracy:±0.05°C (with fixed point cells)
Display Resolution:0.01°C (30-99.99°C); 0.1°C (100-550°C)
Cavity Size:45 mm diameter x 285 mm depth
Heating Time:45 minutes to 550°C
Stabilization Time:45 minutes
Controller Stability (5 min):±0.2°C
Cavity Uniformity:±1°C (100-500°C at 48mm target); ±4°C along 200mm length
Fixed Point Compatibility:Indium (156.5985°C), Tin (231.928°C), Zinc (419.527°C); large fixed point cells available
Power Supply:100-130V or 208-240V AC, 50/60 Hz
Power Consumption:1000 W typical
Dimensions (HxWxD):480 mm x 425 mm x 260 mm
Weight:17 kg
PC Interface:Included with Windows software
Calibration Certificate:UKAS or NPL traceable calibration available
Included Probe:935-14-95 PRT (optional)
Optional Fixed Point Cells:998-06-00A (Indium), 998-06-00B (Tin), 998-06-00C (Zinc), 984-00-00 Gas Flow System