12 month warranty Request a Quote Technical Support
US

Product Specifications

Contact Us

Product Description

The Fluke 4180/4181 Series precision infrared calibrators are designed for calibrating infrared thermometers and thermal imagers. The 4180 covers -15°C to 120°C while the 4181 covers 35°C to 500°C, providing solutions for a wide range of IR temperature measurement devices.

The 152 mm (6-inch) target accommodates the field-of-view requirements of most common infrared thermometers that cannot be accurately tested with smaller targets. Uniformity across the target ensures that variations cannot be detected with a thermal imager at the proper calibration distance.

With accuracy as good as ±0.35°C, the 4180/4181 Series meets its specifications without additional emissivity-related corrections, achieving test uncertainty ratios (TUR) of 4:1. The target has a nominal emissivity of 0.95 and is radiometrically calibrated to minimize emissivity-related uncertainties.

The LCD display continuously shows current temperature and includes a safety feature indicating when the target exceeds 50°C. The indicator remains illuminated when powered and above 50°C, and flashes after power removal until the target cools below 50°C.

Heating time for the 4181 is 45 minutes to 500°C, with stabilization in 10 minutes. The unit includes RS-232 interface and comes with an NVLAP-accredited radiometric calibration report, target cover, user guide, and serial cable.

Features

  • Two models cover -15°C to 500°C total range
  • Display accuracy up to ±0.35°C
  • Stability up to ±0.05°C
  • Uniformity up to ±0.10°C within 2-inch target area
  • 152 mm (6 inch) target accommodates most thermometers
  • Radiometrically calibrated for traceable results
  • Safety indicator for temperatures above 50°C
  • Rapid heating and cooling
  • RS-232 interface included
  • NVLAP accredited calibration report included
  • Lightweight and portable: 9.5 kg
  • Thermometer emissivity compensation range: 0.9 to 1.0
Fluke Calibration

Precision IR Calibrator - -15 to 500°C, 152mm Target | Fluke 4180/4181

Accredited radiometric calibration, 152mm target for most th
MPN/Series: 4180/4181

Product Specifications

Manufacturer:Fluke Calibration (formerly Hart Scientific)

Model Series:4180/4181

Type:Precision Infrared Calibrator

4180 Temperature Range:-15°C to 120°C

4181 Temperature Range:35°C to 500°C

Display Accuracy (4181):±0.35°C at 35°C; ±0.50°C at 100°C; ±0.70°C at 200°C; ±1.20°C at 350°C; ±1.60°C at 500°C

Stability (4181):±0.05°C at 35°C; ±0.20°C at 200°C; ±0.40°C at 500°C

Uniformity (2.0" diameter, 4181):±0.10°C at 35°C; ±0.25°C at 250°C; ±0.50°C at 500°C

Target Diameter:152.4 mm (6 inches)

Nominal Emissivity:0.95 (radiometrically calibrated)

Thermometer Emissivity Compensation Range:0.9 to 1.0

Heating Time (4181):45 minutes (35°C to 500°C)

Cooling Time (4181):100 minutes (500°C to 35°C)

Stabilization Time:10 minutes

Computer Interface:RS-232

Power (4181):115 VAC or 230 VAC, 1000W

Dimensions (HxWxD):356 x 241 x 216 mm (14 x 9.5 x 8.5 in)

Weight (4181):9.5 kg (21 lb)

Included Accessories:NVLAP accredited calibration report, power cord, user guide, documentation CD, target cover, serial cable

Documents: Product Manual , Product Specs

Contact Us
Contact Us